# Radon removal commissioning of the PandaX\-4T cryogenic distillation system

**INSPIRE:** [2743114](https://inspirehep.net/literature/2743114)
**arXiv:** [2401\.01638](https://arxiv.org/abs/2401.01638)
**DOI:** [10\.1088/1748\-0221/19/07/P07010](https://doi.org/10.1088/1748-0221/19/07/P07010)

**Authors:** Cui, Xiangyi, Wang, Zhou, Li, Jiafu, Li, Shuaijie, Si, Lin, Ju, Yonglin, Ma, Wenbo, Liu, Jianglai, Zhao, Li, Ji, Xiangdong, Yan, Rui, Sha, Haidong, Huang, Peiyao, Wang, Xiuli, Liu, Huaxuan

**Submitted:** 3 January 2024

**Subjects:**
- physics\.ins\-det
- Instrumentation

**Journal reference:** JINST 19 \(07\) P07010 \(2024\)

**Citations:** 7

## Abstract

The PandaX\-4T distillation system, designed for the removalof krypton and radon from xenon, is evaluated for its radon removalefficiency using a $\^\{222\}$Rn source during the online distillationprocess\. The PandaX\-4T dark matter detector is employed to monitorthe temporal evolution of radon activity\. To determine the radonreduction factor, the experimental data of radon atoms introducedinto and bypassed the distillation system is compared\. The resultsindicate that the PandaX\-4T distillation system achieves a radonreduction factor exceeding 190 at the flow rate of 10 slpm and thereflux ratio of 1\.44\. Gas\-only online distillation process of a flowrate of 20 slpm is also conducted without observing significantreduction of radon levels in the detector\. This observation suggeststhat the migration flow of radon atoms from the liquid phase to thegas phase is limited, and the flow rate of gas circulation andduration of the process are insignificant compared to the totalxenon mass of 5\.6 tons in the detector\. This study provides theexperimental data to support the efficient removal of radon at ∼Bq level using the PandaX\-4T distillation system, which is theprerequisite of the radon background control in the detector\. Thefurther operation with higher flow rate will be applied for theupcoming science run in PandaX\-4T\.